Evaluation of the accuracy and precision of the microlab 300 device for total cholesterol test on the CHOD-PAP method

Authors

  • Fanny Rizki Ananda Department of Medical Laboratory Technology, Politeknik Kesehatan Kementerian Kesehatan Pontianak, Pontianak, West Kalimantan, Indonesia
  • Hendra Budi Sungkawa Department of Medical Laboratory Technology, Politeknik Kesehatan Kementerian Kesehatan Pontianak, Pontianak, West Kalimantan, Indonesia
  • Laila Kamilla Department of Medical Laboratory Technology, Politeknik Kesehatan Kementerian Kesehatan Pontianak, Pontianak, West Kalimantan, Indonesia
  • Imma Fatayati Department of Medical Laboratory Technology, Politeknik Kesehatan Kementerian Kesehatan Pontianak, Pontianak, West Kalimantan, Indonesia

DOI:

https://doi.org/10.53770/electron.v7i1.516

Keywords:

Quality assurance, Accuracy, Precision

Abstract

Internal quality assurance (PMI) is a routine preventive and monitoring activity in the laboratory to minimize errors and produce accurate tests. Quality accuracy and precision are important indicators in assessing PMI. Accuracy indicates the closeness of the test results to the true value, while precision describes the closeness of the results of repeated tests on the same sample. Total cholesterol examination using a spectrophotometer, especially the CHOD-PAP method, is often carried out at the Clinical Laboratory of the Poltekkes Kemenkes Pontianak due to its stability. This observational study aims to evaluate the accuracy and precision of the Microlab 300 tool for total cholesterol examination using the CHOD-PAP method at the Integrated Laboratory of the Poltekkes Kemenkes Pontianak. The study population was the assayed control serum solution, with samples in the form of normal control serum solution dialab which was divided into 100 vials containing 50 µL. A total of 30 vials were used for the preliminary period and 30 vials for the control period. The parameter examined was total cholesterol by CHOD-PAP method. The results showed the average accuracy of the Microlab 300 tool was 96.08% and precision was 99.21%. The daily graph using Westgard's rule shows that the control material is within the control limits. Thus, it can be concluded that the Microlab 300 tool shows good accuracy and precision in the examination of total cholesterol CHOD-PAP method in the laboratory

References

Alemnji, G. A., Zeh, C., Yao, K., & Fonjungo, P. N. (2014). Strengthening national health laboratories in sub‐S aharan A frica: a decade of remarkable progress. Tropical medicine & international health, 19(4), 450-458.https://doi.org/10.1111/tmi.12269

Cheng, W. L., Markus, C., Lim, C. Y., Tan, R. Z., Sethi, S. K., & Loh, T. P. (2023). Calibration practices in clinical mass spectrometry: review and recommendations. Annals of laboratory medicine, 43(1), 5-18. https://doi.org/10.3343/alm.2023.43.1.5

Chackartchi, T., Romanos, G. E., Parkanyi, L., Schwarz, F., & Sculean, A. (2022). Reducing errors in guided implant surgery to optimize treatment outcomes. Periodontology 2000, 88(1), 64-72.https://doi.org/10.1111/prd.12411

Del Olmo, A., De Lacalle, L. L., De Pissón, G. M., Pérez-Salinas, C., Ealo, J. A., Sastoque, L., & Fernandes, M. H. (2022). Tool wear monitoring of high-speed broaching process with carbide tools to reduce production errors. Mechanical Systems and Signal Processing, 172, 109003.https://doi.org/10.1016/j.ymssp.2022.109003

Lebo, T., Merkel, P. A., & Knight, V. (2024). Validation and Quality Control: General Principles and Application to the Clinical Immunology Laboratory. Manual of Molecular and Clinical Laboratory Immunology, 1, 369-383. https://doi.org/10.1002/9781683674023.ch35

McPherson, R. A., & Pincus, M. R. (2021). Henry's clinical diagnosis and management by laboratory methods E-book. Elsevier Health Sciences.

Milne, K. C., & Milne, T. L. (2010). Public health laboratory system improvement program: development and implementation. Public Health Reports, 125(2_suppl), 31-39.https://doi.org/10.1177/00333549101250S204

Nabilla, J., Ulya, N. I., Rofikoh, R., Agiska, S. D., Nadifah, F., & Qomariyah, N. (2023). Validasi Pemeriksaan Kolesterol Total Metode CHOD-PAP pada Sampel yang Disimpan Selama 1 Minggu dalam Suhu 4-8 C CHOD-PAP Method Total Cholesterol Examination Validation on Samples Stored for 1 Week In 4-8 C Temperature. Jurnal Laboratorium Medis, 5(1), 7-11.

Olver, P., Bohn, M. K., & Adeli, K. (2023). Central role of laboratory medicine in public health and patient care. Clinical Chemistry and Laboratory Medicine (CCLM), 61(4), 666-673.

Pamungkas, G. C., Handayati, A., & Woelansari, E. D. (2019). Gambaran Pemantapan Mutu Eksternal Laboratorium Parameter Eritrosit Dan Trombosit Di Puskesmas Wilayah Kabupaten Mojokerto. Analis Kesehatan Sains 8(2), 704–709.

Siregar, M. H., Susanti, R., Indriawati, R., Panma, Y., Hanaruddin, D. Y., Adhiwijaya, A., ... & Renaldi, R. (2022). Metodologi penelitian kesehatan. Yayasan Penerbit Muhammad Zaini.

Sudaryono, D. (2017). Metodologi Penelitian: Penelitian Kualitatif. Depok: PT. Rajagrafindo Persada.

Sujarweni, V. W. (2014). Metodelogi Penelitian. Yogyakarta: Pustaka Baru Perss.

Wijayanti, H. N., Yansen, A., & Nurcahyanti, O. (2022). Analisis Faktor-Faktor Yang Mempengaruhi Pemantapan Mutu Internal Pada Pemeriksaan Urin Rutin Di Laboratorium Rsud Kota Depok. Binawan Student Journal, 4(1), 38-42.

Swanson, K., Dodd, M. R., VanNess, R., & Crossey, M. (2018). Improving the delivery of healthcare through clinical diagnostic insights: a valuation of laboratory medicine through “Clinical Lab 2.0”. The Journal of Applied Laboratory Medicine, 3(3), 487-497.https://doi.org/10.1373/jalm.2017.025379

Downloads

Published

2025-01-31

How to Cite

Ananda, F. R., Sungkawa, H. B., Kamilla, L., & Fatayati, I. (2025). Evaluation of the accuracy and precision of the microlab 300 device for total cholesterol test on the CHOD-PAP method. ELECTRON (Journal of Science and Technology), 7(1), 17–26. https://doi.org/10.53770/electron.v7i1.516